Description
Waygate Technologies Krautkramer DMS Go+ Ultrasonic Thickness Gauges, a Baker Hughes business product (formerly GE Inspection Technologies), provide accurate, reliable, and comprehensive thickness inspection data in a wide range of applications and environments. The DMS GO+ features the same hardware and operating platform as the USM Go+ flaw detector, combining the two to offer a powerful instrument. These Gauges are lightweight, ergonomic, and IP67 rated for operating in harsh environments. The DMS Go+ is ideal for inspections in oil & gas, power generation, aerospace, and metals industries.
The DMS Go+ includes automatic gain control for repeatability and corrosion monitoring, and built-in temperature compensation for accurate measurement up to 1000°F. An on-board data recorder can hold thousands of thickness readings including A-scan, B-scan, and MicroGrid attachments, which can be exported to user data management systems in various file formats. Three models are available to best suit a wide variety of inspection needs.
DMS Go+ Basic
- Simple to use
- A-scan verification
- D-Meter dual-element probe compatible
DMS Go+ Standard
- B-scan, timed, for cross-sectional view
- Compatible with single element probes for precision thickness measurement and temperature compensated readings
DMS Go+ Advanced
- 6 data recording file structures plus 3D & 4D via UltraMATE
- TopCoat/Auto-V measurement capabilities to eliminate removal of many coatings
Технические характеристики
Specs
Display | 5 800 x 480 pixels (W x H) 108 65 mm / 4.25 x 2.55 >,400 cd/m2 |
Size (WxHxD) | 175 x 111 x 50 mm / 6.88 x 4.37 x 1.96 |
Weight | 877 g / 1.87 lbs |
Protection class | IP 67 |
Operating temperature | 0 55 C / 32 131 F |
Battery | Li-Ion, rechargeable >, 8 hours operation time |
Power adapter/charger | 100 240 V AC, 50/60 Hz |
Probe connector | Dual Lemo-00 (T/R) |
PC interface | Micro USB |
Memory card | SD-Card 16 GB max |
Data recorder (model specific) | 100.000 readings per file. Multiple files can be stored on SD card. 8 file formats. Attachment of A-Scan, B-Scan and micro grid |
Pulser | 120 300 V Spike wave Automatically matched to probe |
Pulse Repetition Frequency | 4, 8 or 16 Hz selectable |
Receiver (model specific) | 110 dB dynamic. Automatic gain control. Manual high, -low, -auto |
Measurement range | 0.4 14,000 m (0.01 551) |
Units | mm, inch, s |
Digital display resolution | 0,01 mm or 0,1 mm (0.001 or 0.01) selectable |
Measurement techniques (model specific) | Zero crossing. IP to 1st echo. Multi echo, TopCoat, Auto-V |
Calibration (model specific) | One-point. Two-point. Auto or Manual On-block and Off-block Zero. Automatic V-Path correction |
Display mode (model specific) | Thickness and A-Scan. Temperature corrected thickness. B-Scan. Min/Max capture. Differential |
Compliance | EN 61010, EN 61326-1, EN 12668, ASTM E 1324, E317, ANSI/ NCSL Z 540-1-1994, MIL-STD 45662A, MIL-STD 2154, EN 15317 |