CytoFLEX V2-B3-R2 Flow Cytometer has 3 active lasers (405 nm, 488 nm, 638 nm) and 7 channels for fluorescence detection. This instrument can be upgraded with a maximum of 5 Violet, 5 Blue, and 3 Red channels. It includes 13 repositionable bandpass filters, 450/45, 525/40 (2), 585/42, 610/20 (2), 660/10 (2), 690/50, 712/25, 780/60 (3). The complete system is supplied with computer work station with CytExpert Acquisition software, sheath and waste containers with harness.
Технические характеристики
Product Category (BU) | Flow Cytometry Analyzer |
Software | CytExpert software |
Regulatory Status | RUO |
Resolution | <,300 nm particles (488 nm side scatter), <,80 nm polystyrene particles (405 nm side scatter), rCV 3% |
Performance | <,30 MESF for FITC, <,10 MESF for PE |
Number of Parameters | 7 fluorescence detectors |
Optical Filters | 450/45, 525/40 (2), 585/42, 610/20 (2), 660/10 (2), 690/50, 712/25, 780/60 (3) |
Detectors | Photodiodes |
Flow Cell | Patent-pending, alignment-free integrated optics quartz flow cell (420 μm x 180 μm ID) design with >,1.3 numerical aperture |
Signal Processing | Digital system with 7-decade display |
Power Requirements | 100-240 VAC, 50/60 Hz |
Weight | 23.4 kg(51.6 lb) |
Width | 42.5 cm(16.7 in) |
Depth | 42.5 cm(16.7 in) |
Height | 34 cm(13.4 in) |