CytoFLEX S V2-B2-Y3-R2 Flow Cytometer has 4 active lasers (405 nm, 488 nm, 561 nm, 638 nm) and 9 channels for fluorescence detection. This instrument can be upgraded with a maximum of 4 Violet, 2 Blue, 4 Yellow Green, and 3 Red channels. It includes 13 repositionable bandpass filters, 450/45, 525/40 (2), 585/42, 610/20 (2), 660/10 (2), 690/50 (2), 712/25, 780/60 (2). The complete system is supplied with computer work station with CytExpert Acquisition software, sheath and waste containers with harness.
Технические характеристики
Product Category (BU) | Flow Cytometry Analyzer |
Software | CXP software |
Regulatory Status | RUO |
Resolution | 0.5 μm diameter particles. HPCV is <,2% to <,4% depending on the laser. |
Performance | <,600 MESF for APC, <,600 MESF for FITC, <,300 MESF for PE |
Number of Parameters | {7061A7CA-57B5-40B8-A45D-F31B7F66A7DC} |
Optical Filters | 450/45, 525/40 (2), 585/42, 610/20 (2), 660/10 (2), 690/50 (2), 712/25, 780/60 (2) |
Detectors | Photodiodes (forward & side scatter), Photomultiplier tubes (fluorescence) |
Flow Cell | BioSense 150 μm x 450 rectangular channel with an integral lens |
Signal Processing | Digital System with 4 decade data display |
Power Requirements | 100-240V, 50-60Hz, Requires dedicated lines |
Weight | 84.8 kg(187 lb) |
Width | 97.8 cm(38.5 in) |
Depth | 88.9 cm(35 in) |
Height | 61 cm(24 in) |