Product Description 200-LITER CAPACITY | TEMP, CO2, HUMIDITY, & O2 CONTROL | HEPA FILTRATION | STERILIZATION CYCLES This line of CO2 incubators offers touch panel control and monitoring of temperature, CO2, relative humidity, and oxygen levels to meet the demands of a broad range of cell culture applications. Beyond using an infrared sensor for precise control of CO2 and separately adjustable chamber, door and perimeter heaters to balance internal temperature, this family can create hypoxic conditions by injecting N2 and directly control humidity to set point by routing air through an evaporator tank. The air pump outside the 200L growth chamber also constantly draws air through a HEPA filter and cycles it back to create ISO Class 5 quality air at positive pressure and minimize contamination. Scientists can choose between a 95C humidified or 145C dry decontamination depending on their laboratorys procedures. A touch screen makes it easy to enter set points as well as to view or download historical performance. This moderately sized incubator can be stacked to efficiently use the valuable space in your laboratory. Quick Specs AIR CLEANLINESS (ISO 14644) ISO-Class-5 CO2 ACCURACY &plusmn, 0.1% CO2 RECOVERY Up to 5% &plusmn, 0.2% / -0.5 in 5 minutes Average CONTROL SYSTEM NuTouch TEMPERATURE ACCURACY C &plusmn, 0.1% TEMPERATURE CONTROL Direct Heat TEMPERATURE RECOVERY 0.12C/min. on Average TEMPERATURE UNIFORMITY C &plusmn, 0.35C @ 37C Features Cold Rolled Steel with Powder Coated Finish Exterior Polished, 16 Gauge Type 304L Stainless-Steel Chamber Set of 4 Stainless Steel Shelves Inner Glass Door Separately Adjustable Perimeter & Door Heaters Infrared CO2 Sensor Dual Temperature Sensors Air Pump External to Chamber HEPA Filter External to Chamber Multilingual Touch Screen USB Port for Data Download RH Sensor and Evaporator Tank #col-1795915675 > .col-inner { margin: 0px 0px -49px 0px, } #gap-254309674 { padding-top: 36px, } THE MODERN USER EXPERIENCE YOU EXPECT. NUTOUCH CONTROL SYSTEM The NuTouch control system provides you with the user experience you have come to expect from any piece of modern equipment. Change set points, view chamber history, start a decontamination cycle, or perform a calibration procedure from this simple interface. All NuAire CO2 incubators standardize on the NuTouch control system, meaning that once users learn to navigate one incubator they can move seamlessly to any other in the lab. THE WARM EMBRACE OF DIRECT HEAT TECHNOLOGY. DIRECTLY HEATED CHAMBER The direct heat incubator uses heating elements placed around the chamber, in the door perimeter, and in the door itself to provide balanced temperature within a steel, insulated chamber environment. Door and perimeter heaters can be adjusted independently to ensure uniformity even when the incubator is installed in more demanding ambient temperatures. #col-1359113752 > .col-inner { padding: 0px 30px 0px 0px, margin: 0px 0px -13px 0px, } CONVENIENT CONTAMINATION CONTROL. 145C HEATED DECONTAMINATION The direct heat technology enables the use of a 145C decontamination cycle that can be conveniently run overnight in just 8 hours without removing the CO2, temperature, or O2 sensors. The NuTouch control system steps the user through the process with on screen instructions and displays progress throughout each phase of the complete cycle. CELL CULTURES DESERVE THEIR OWN CLEAN ROOM. CONSTANT HEPA AIR FILTRATION Continual HEPA filtration of chamber air complements periodic high heat decontamination to provide Constant Contamination Control. InVitrocells design constantly takes a sample of air from the chamber, routes it through a HEPA filter in a compartment outside that inner chamber, and ultimately returns it to create an ISO Class 5 level of clean air in the interior as well as a slight positive pressure. Inside the InVitrocell, your cell cultures are protected by a multi-tiered defense against contamination. #col-1281435814 > .col-inner { padding: 0px 0px 0px 30px, margin: 0px 0px -7px 0px, } Specifications Est. Shipping Width Inches (mm) 31 (787) Est. Shipping Height Inches (mm) 51(1219) Est. Shipping Depth Inches (mm) 31 (787) Access Port Location Right Side Air Cleanliness (ISO 14644) ISO-Class-5 Chamber Air Circulation Method Air Pump Chamber Air Filter 99.97% Efficiency @ &ge, 0.3 Micron Capsule Filter Located in Sensor Bay Chamber Heater(s) Location Walls, Top and Bottom of Chamber Chamber Jacket Technology Air Jacket via One or Two Axial Fans Drawing Air into Shell CO2 Accuracy &plusmn, 0.1% CO2 Control Logic Fixed Algorithm/Manual, Environmental Adaptable CO2 Display Resolution 0.1% CO2 Range 0.1 to 20%, 5% Default Set Point CO2 Recovery Up to 5% &plusmn, 0.2% / -0.5 in 5 minutes Average CO2 Sensor Type Infrared (NDIR) Control System NuTouch Data Output Features RS-485, USB, 4-20mA Door and Perimeter Heater Control Logic Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable Electrical Configuration 115V50Hz, 230V5060Hz Exterior Depth Inches (mm) 27.188 (691) Exterior Door Seal Magnetic Outer Door Gasket Exterior Height Inches (mm) 39.688 (1008) Exterior Material Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish Exterior Width Inches (mm) 26.750 (680) Footprint Depth inches (mm) 20.75 (527) Footprint Width inches (mm) 23.75 (603) In-line Gas Filters 99.97% Efficiency @ &ge, 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor Segmented Doors Optional Set of 6 Inner Doors, 2 Across by 3 Down Inner Doors One Insulation High Density Insulation with High R Rating Interior Depth Inches (mm) 20.053 (509) Interior Door Seal Silicone Rubber Gasket Interior Height Inches (mm) 28.500 (724) Interior Material Polished Type 304L 16 Gauge Stainless Steel Interior Volume Cubic Feet (Liters) 7.06 (200) Interior Width Inches (mm) 21.257 (540) Net Weight lbs (kg) 235 (106) O2 Accuracy &plusmn, 0.25% O2 Control Logic Fixed algorithm/ manual environmental adaptable O2 Display Resolution 0.1% O2 Measuring Range 0 25% O2 Set Point Range 0.5 21.0% (ambient) (21.0% Default) (2 21.0% when RH setpoint is above 90%) O2 Recovery 5.0 &plusmn, 0.5% twenty minutes on Average O2 Sensor Type Zirconia Ceramic Power Cord Length Foot (Meter) 8 (2.5) Remote Alarm Terminals Standard RH Accuracy +5% /- 3% RH Control Method Water Pan or Evaporator Box, Depending on Model RH Control Logic Fixed algorithm/ manual environmental adaptable RH Control Range Ambient dependent to 95% RH Display Resolution 0.1% RH Evaporator Box Capacity 3L RH Recovery 90% +5%/-3% within 25 minutes on average RH Sensor Type Precision capacitance bulk-polymer RH Water Pan Capacity 2L (Recommended Fill 1.5L) Sample Port Location Inner Glass Door Shelf Capacity lbs (kg) 25 lbs per Shelf / 125 lbs per Incubator Shelf Depth Inches (mm) 18 3/4 (476) Shelf Material Stainless Steel Shelf Width Inches (mm) 19 3/4 (502) Shelves 4 Shelves Supplied, 20 Shelves Maximum Capacity Est. Shipping Weight lbs (kg) Estimated 325 lbs (147 kgs) Decontamination Cycle True (False on NU-5800, NU-5800E) Temperature Accuracy C &plusmn, 0.1% Temperature Control Direct Heat Temperature Control Range 5C above ambient (to a 30C max ambient) to 55C Temperature Display Resolution 0.1C Temperature Recovery 0.12C/min. on Average Temperature Sensor Type Precision Integrated Circuit Temperature Set-Point Range 5C to 55C (37.0C Default) Temperature Uniformity C &plusmn, 0.35C @ 37C Water Pan Depth 1.5 (38) Water Pan Length Inches (mm) 12 (305) Water Pan Width 10 (254)

Технические характеристики

Technical Specifications

Est. Shipping Width Inches (mm)31 (787)
Est. Shipping Height Inches (mm)51(1219)
Est. Shipping Depth Inches (mm)31 (787)
Access Port LocationRight Side
Air Cleanliness (ISO 14644)ISO-Class-5
Chamber Air Circulation MethodAir Pump
Chamber Air Filter99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay
Chamber Heater(s) LocationWalls, Top and Bottom of Chamber
Chamber Jacket TechnologyAir Jacket via One or Two Axial Fans Drawing Air into Shell
CO2 Accuracy± 0.1%
CO2 Control LogicFixed Algorithm/Manual, Environmental Adaptable
CO2 Display Resolution0.1%
CO2 Range0.1 to 20%, 5% Default Set Point
CO2 RecoveryUp to 5% ± 0.2% / -0.5 in 5 minutes Average
CO2 Sensor TypeInfrared (NDIR)
Control SystemNuTouch
Data Output FeaturesRS-485, USB, 4-20mA
Door and Perimeter Heater Control LogicBase Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable
Electrical Configuration115V50Hz, 230V5060Hz
Exterior Depth Inches (mm)27.188 (691)
Exterior Door SealMagnetic Outer Door Gasket
Exterior Height Inches (mm)39.688 (1008)
Exterior MaterialCold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish
Exterior Width Inches (mm)26.750 (680)
Footprint Depth inches (mm)20.75 (527)
Footprint Width inches (mm)23.75 (603)
In-line Gas Filters99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor
Segmented DoorsOptional Set of 6 Inner Doors, 2 Across by 3 Down
Inner DoorsOne
InsulationHigh Density Insulation with High ‘R’ Rating
Interior Depth Inches (mm)20.053 (509)
Interior Door SealSilicone Rubber Gasket
Interior Height Inches (mm)28.500 (724)
Interior MaterialPolished Type 304L 16 Gauge Stainless Steel
Interior Volume Cubic Feet (Liters)7.06 (200)
Interior Width Inches (mm)21.257 (540)
Net Weight lbs (kg)235 (106)
O2 Accuracy± 0.25%
O2 Control LogicFixed algorithm/ manual environmental adaptable
O2 Display Resolution0.1%
O2 Measuring Range0 — 25%
O2 Set Point Range0.5 — 21.0% (ambient) (21.0% Default) (2 — 21.0% when RH setpoint is above 90%)
O2 Recovery5.0 ± 0.5% twenty minutes on Average
O2 Sensor TypeZirconia Ceramic
Power Cord Length Foot (Meter)8 (2.5)
Remote Alarm TerminalsStandard
RH Accuracy+5% /- 3%
RH Control MethodWater Pan or Evaporator Box, Depending on Model
RH Control LogicFixed algorithm/ manual environmental adaptable
RH Control RangeAmbient dependent to 95%
RH Display Resolution0.1%
RH Evaporator Box Capacity3L
RH Recovery90% +5%/-3% within 25 minutes on average
RH Sensor TypePrecision capacitance bulk-polymer
RH Water Pan Capacity2L (Recommended Fill 1.5L)
Sample Port LocationInner Glass Door
Shelf Capacity lbs (kg)25 lbs per Shelf / 125 lbs per Incubator
Shelf Depth Inches (mm)18 3/4 (476)
Shelf MaterialStainless Steel
Shelf Width Inches (mm)19 3/4 (502)
Shelves4 Shelves Supplied, 20 Shelves Maximum Capacity
Est. Shipping Weight lbs (kg)Estimated 325 lbs (147 kgs)
Decontamination CycleTrue (False on NU-5800, NU-5800E)
Temperature Accuracy °C± 0.1%
Temperature ControlDirect Heat
Temperature Control Range5°C above ambient (to a 30°C max ambient) to 55°C
Temperature Display Resolution0.1°C
Temperature Recovery0.12°C/min. on Average
Temperature Sensor TypePrecision Integrated Circuit
Temperature Set-Point Range5°C to 55°C (37.0°C Default)
Temperature Uniformity °C± 0.35°C @ 37°C
Water Pan Depth1.5 (38)
Water Pan Length Inches (mm)12 (305)
Water Pan Width10 (254)