Технические характеристики
The HPD II Aerosol Monitoring Accessory is compatible with non-toxic, non-flammable gases such as Clean Dry Air (CDA), Argon, Nitrogen and Carbon dioxide. Sampling gases at pressures greater than sea level, from 40 up to 150 psi, is challenging for particle counters. Connecting pressurized air or gas to a particle counter can defeat its flow metering system and possibly damage the instrument. High Pressure Diffusers (HPDs) have been developed to address particle counting in compressed gas. HPDs reduce the pressure of the gas to about atmospheric pressure by allowing some of the air/gas to diffuse out of the system and allow the Lasair III particle counter’s flow-metering system to function as designed. Specifically designed for use with the Lasair III-110 particle counter • Simple connections for gas inlet/outlet • Filtration of any excess gas • Particle counter can remain on, even when gas supply is off Two models to reduce high-pressure gas between 40 – 100 psi or 75 – 150 psi to about atmospheric pressure Use in conjunction with a Lasair III-110 particle counter to measure particles as small as 0.1 µm in CDA (compressed dry air), nitrogen, carbon dioxide or argon Allows users to efficiently monitor pressurized gases with standard aerosol particle counters