Product Description Refurbished Midmark Ritter M11D Ultraclave, Previous Model WARRANTY: 6 month parts and labor . Sterilization is easy with the Midmark M11D Autoclave. Just a few simple steps completes the process. Simply press the Start/Standby button and the Ritter M11 automatically fills with the correct amount of water. Select the cycle you want and the time is set for you. Cycle time from start to finish is 28 minutes (if cold) and 18 minutes (if hot) for unwrapped cycles. Features Low water indicator light tells you when the reservoir level is too low A door-ajar light tells you if the door is not securely closed LED display shows the units temperature during heat-up and time remaining during cycles. Lights turn on progressively during filling, heating, sterilizing, drying and complete stages. Easy to fill and to drain water located at the front of the machine A large capacity water reservoir means you fill it less often! Four stainless steel trays come with the unit (two large, two small) Easy-to-read pressure gauge and electronic controls offer simple operation and reduce down time Technical Specifications Chamber Size: 11 x 18 Volume: 6.5 gal (24.6 L) Dimensions (L x W x H): 23.8 x 17.8 x 17.8 Trays: 4 Tray Size: Two Large: 9W x 15L x 1.1D & Two Small: 6.6W x 15L x 1.1D Voltage: 115V Frequency: 50/60 Current: 15A Shipping Weight: 131 lbs Minimum Countertop Area: 21 D x 17.8 W Sterilization Programs Temp. Hot Cycle Time (fill, heat-up & vent) Sterilization Time Dry Time Total Hot Cycle Time (without dry) Total Hot Cycle Time (with dry) Description Icon Unwrapped 270 F 15 min 3 min 30 min 18 min 48 min Pouches 270 F 17 min 5 min 30 min 22 min 52 min Packs 250 F 14 min 30 min 30 min 44 min 74 min Handpieces 270 F 16 min 6 min 30 min 22 min 52 min
Технические характеристики
Technical Specifications
- Chamber Size: 11′ x 18′
- Volume: 6.5 gal (24.6 L)
- Dimensions (L x W x H): 23.8′ x 17.8′ x 17.8′
- Trays: 4
- Tray Size: Two Large: 9’W x 15’L x 1.1’D & Two Small: 6.6’W x 15’L x 1.1’D
- Voltage: 115V
- Frequency: 50/60
- Current: 15A
- Shipping Weight: 131 lbs
- Minimum Countertop Area: 21′ D x 17.8′ W
Sterilization Programs | Temp. | Hot Cycle Time (fill, heat-up & vent) | Sterilization Time | Dry Time | Total Hot Cycle Time (without dry) | Total Hot Cycle Time (with dry) | |
Description | Icon | ||||||
Unwrapped | 270° F | 15 min | 3 min | 30 min | 18 min | 48 min | |
Pouches | 270° F | 17 min | 5 min | 30 min | 22 min | 52 min | |
Packs | 250° F | 14 min | 30 min | 30 min | 44 min | 74 min | |
Handpieces | 270° F | 16 min | 6 min | 30 min | 22 min | 52 min |