Поставка оптоэлектронных компонентов ведущих мировых производителей

Амерекс HRH-110

Доставка импортных компонентов Amerex Instruments по России от 3х недель транспортными компаниями СДЭК, Деловые Линии, Major Express.
Оплата физическими и юридическими лицами безналичным расчетом.
Цена может измениться после запроса у поставщика!
Артикул: 8021-10-0014 TME арт.: 8021-10-0014

Product Description Front-loading design Fully automatic operation from start to finish with or without drying. One-touch door closing lever opens and closes the door easily, surely and safely. Pre-vacuum cycle vacuum pump removes air from the chamber when sterilizing solids such as glassware and pipette tips, thereby eliminating cold spots, and ensuring faster and more efficient sterilization. Drying by heat and vacuum increases drying efficiency and reduces cycle times. Additional drying mode for complete drying without having to run the sterilization cycle again. Bacteria filter reduces re-contamination. Lid safety interlock prevents lid from opening during a cycle, when the chamber is pressurized, or when the chamber temperature is above 97C at which super-heated liquids might boil over when disturbed. Door-closure sensor prevents a cycle from starting unless the door is properly closed. HRH-110 size: 110 liters, 16.5 diameter, 31.3 depth. Technical Specifications SPECIFICATIONS HRH-110 Self-contained Design Does not require in-house steam, plumbing or a floor drain hole for installation Sterilization Temperature 105C 135C (programmable) Temperature Display Range 5C 137C Sterilization Timer 1 ~ 60 min (remaining time displayed) Drying Timer 1 ~ 120 min (remaining time displayed) Drying Method Vacuum pump and heating Sterilization and Operation Modes For solids pre-vacuum sterilization drying For solids pre-vacuum sterilization fast exhaust For liquids sterilization slow exhaust For drying additional drying after mode #1 end Chamber Water Source Auto water filling from an internal 10-liter water tank Maximum allowable Pressure (Gauge Pressure) 0.206 MPa (2.06 kg/cm2G) Chamber Size 16.5 Diameter x 31.3 depth (420 Diameter x 795 depth) mm Chamber Material Stainless steel SUS304 Outer Dimensions 27.2W x 45.3D x 54.5H in, (690W x 1150D x 1385H) mm Pressure Gauge Analog display: -0.1 ~ 0.4 MPa (-1 to 4 bar) Safety Devices Pressure relief valve, circuit breaker with GFCI, vacuum pump breaker, low-water power cut-off, insufficient water in the exhaust bottle warming lamp, error messages for 1) wire breakage in temperature sensor, 2) over-heating, 3) over-cooling, 4) over-pressure, and 5) door is not properly locked Standard Accessories Drain hose, water supply hose, caster stopper Options 1) Load sensor, square type wire basket, wire basket with solid bottom, dressing case, 2) Data logger, and 3) Digital printer Power Source/Consumption AC220V, 50/60Hz, 1 phase, 3.2kW Net Weight (approx.) 462 lb, (210 kg) Warrenty 1 year

Технические характеристики

Technical Specifications

SPECIFICATIONS HRH-110
Self-contained Design Does not require in-house steam, plumbing or a floor drain hole for installation
Sterilization Temperature 105ºC – 135ºC (programmable)
Temperature Display Range 5ºC — 137ºC
Sterilization Timer 1 ~ 60 min (remaining time displayed)
Drying Timer 1 ~ 120 min (remaining time displayed)
Drying Method Vacuum pump and heating
Sterilization and Operation Modes For solids – pre-vacuum → sterilization → drying For solids – pre-vacuum → sterilization → fast exhaust For liquids – sterilization → slow exhaust For drying – additional drying after mode #1 → end
Chamber Water Source Auto water filling from an internal 10-liter water tank
Maximum allowable Pressure (Gauge Pressure) 0.206 MPa (2.06 kg/cm2G)
Chamber Size 16.5′ Diameter x 31.3′ depth (420 Diameter x 795 depth) mm
Chamber Material Stainless steel SUS304
Outer Dimensions 27.2W x 45.3D x 54.5H in, (690W x 1150D x 1385H) mm
Pressure Gauge Analog display: -0.1 ~ 0.4 MPa (-1 to 4 bar)
Safety DevicesPressure relief valve, circuit breaker with GFCI, vacuum pump breaker, low-water power cut-off, insufficient water in the exhaust bottle warming lamp, error messages for 1) wire breakage in temperature sensor, 2) over-heating, 3) over-cooling, 4) over-pressure, and 5) door is not properly locked
Standard Accessories Drain hose, water supply hose, caster stopper
Options 1) Load sensor, square type wire basket, wire basket with solid bottom, dressing case, 2) Data logger, and 3) Digital printer
Power Source/Consumption AC220V, 50/60Hz, 1 phase, 3.2kW
Net Weight (approx.) 462 lb, (210 kg)
Warrenty 1 year