The 819D-SL-3.3-CAL2 (P/N 7N6320A) is a NIST traceable calibrated integrating sphere detector using a 3.3 inch integrating sphere with a high performance silicon photodetector. The input port size is 0.5 in. diameter and the wavelength range is 400 – 1100 nm. It comes with an SMA fiber optic connector on the North pole as a standard feature allowing a small amount of light pickoff for wavelength measurement or any further analysis without affecting the overall system calibration. Note that the system calibration is no longer valid if any component is changed from the original calibrated configuration. Calibration is done only with no attenuator on the photodiode.
Технические характеристики
- Type Sphere Sensor
- Detector Type Diverging Beam Integrating Sphere
- Detector Input Free Space
- Fiber-Optic Connector SMA on the north pole
- Detector Material Silicon
- Spectral Range 400 — 1100 nm
- Sensor Size Ø11.3 mm
- Attenuator OD1, Built-in
- Maximum Measurable Power 1650 mW
- Minimum Measurable Power 100 nW
- Maximum Measurable Power Benchtop Meter 1.65 W
- Maximum Measurable Power Handheld Meter 1.5 W
- Maximum Measurable Power Without Attenuator 165 mW (Benchtop)
150 mW (Handheld) - Calibration Module DB15 Connector Pigtail
- Calibration Uncertainty 1.65% (400-430 nm)
1.1% (430-1000 nm)
4.3% (1035-1065 nm) - Linearity ±1 %
- Sphere Size 3.3 in.
- Sphere Material PTFE
- Input Port Size 0.5 in.
- Connector Type DB15
- Cable Length 2 m